Core Optical Specifications
• Measurement Principle: Dual-channel scattered light (forward scattering) paired with a direct reference beam.
• Scattering Angle: 11°.
• Detectors: 8 hermetically encapsulated silicon photodiodes for scattered light, and 1 direct-beam photodiode for reference and high-concentration tracking.
• Wavelength Range: 730 – 970 nm (Near-Infrared / NIR).
• Light Source: Incandescent tungsten lamp (Nominal 5.0 V DC).
• Measurement Range: Typically 0 – 0.5 ppm to 500 ppm (equivalent to 0–200 FTU / 0–50 EBC depending on path length).
• Resolution: < ±0.05%
• Repeatability: < ±0.5%
Material & Process Limits
Itjdpfx Ao Aayzieqren
• Sensor Windows: Manufactured from single-crystal sapphire for maximum resistance to chemical corrosion and abrasion.
• Housing Material: Stainless steel (standard 1.4571 / 316Ti).
• Line Sizes Supported: Accommodates process pipes from DN 6 to DN 200 (1/4" to 8") via customized flow cells.
• Process Temperature:
• Standard version: Permanent 0°C to +120°C (with short daily peaks up to 150°C).
• High-temperature (HT) version: -20°C to +240°C.
• Process Pressure: Supports application environments from 10 mbar up to 325 bar depending on the specific line fitting and configuration.
Good working condition. Please check out our other listings as well.
• Measurement Principle: Dual-channel scattered light (forward scattering) paired with a direct reference beam.
• Scattering Angle: 11°.
• Detectors: 8 hermetically encapsulated silicon photodiodes for scattered light, and 1 direct-beam photodiode for reference and high-concentration tracking.
• Wavelength Range: 730 – 970 nm (Near-Infrared / NIR).
• Light Source: Incandescent tungsten lamp (Nominal 5.0 V DC).
• Measurement Range: Typically 0 – 0.5 ppm to 500 ppm (equivalent to 0–200 FTU / 0–50 EBC depending on path length).
• Resolution: < ±0.05%
• Repeatability: < ±0.5%
Material & Process Limits
Itjdpfx Ao Aayzieqren
• Sensor Windows: Manufactured from single-crystal sapphire for maximum resistance to chemical corrosion and abrasion.
• Housing Material: Stainless steel (standard 1.4571 / 316Ti).
• Line Sizes Supported: Accommodates process pipes from DN 6 to DN 200 (1/4" to 8") via customized flow cells.
• Process Temperature:
• Standard version: Permanent 0°C to +120°C (with short daily peaks up to 150°C).
• High-temperature (HT) version: -20°C to +240°C.
• Process Pressure: Supports application environments from 10 mbar up to 325 bar depending on the specific line fitting and configuration.
Good working condition. Please check out our other listings as well.
Aanbieder
Opmerking: Registreer gratis of log in, om alle informatie te bekijken.
Laatst online: Vandaag
Geregistreerd sinds: 2020
271 advertenties online
Aanvraag sturen
Telefoon & Fax
+36 62 99... weergeven
Uw advertentie is succesvol verwijderd
Er is een fout opgetreden